Example: FTIR Spectropolarimetry
for Sample Materials
FTIR in emission mode with P-SIM Module retrofitted to
front of spectrometer
Grazing angle specular reflectance
800
900
1000
1100
1200
Wavenumber @cm-1D
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
FTIR P-SIM spectrum obtained at 75-deg incidence angle, viewing a water
scene filling the FOV. Clear sky background, -4C ambient temperature, 37C
water temperature. The fringes indicate the expected 2-4% polarization.