Long path length - low volume fast response absorption cell for tunable laser spectroscopy and atmospheric trace gas detection

 

AMAC-100

AMAC-200

AMAC-36 AMAC-76
Path length

100 (49) m

210 (153, 295) m

36 (18) m

76 (55)

Base path

0.55 m

0.88 m

0.20 m

0.32 m

Number of passes

182 (90)

238 (174, 334)

182 (90)

238 (174)

Volume

3.2 liters

5.1 liters

0.3 liters

0.5 liters

Mirror diameter

7.6 cm

7.6 cm

3.8 cm

3.8 cm

Optical axis height

8.9 cm

8.9 cm

6.4 cm

6.4 cm

In-out half-angle

3.7 degrees

2.5 degrees

5.1 degrees

3.2 degrees

Input hole diameter

5.3 mm

5.3 mm

4.3 mm

4.3 mm

Dimensions exterior

  Length
  Width
 
Weight




70 cm
13.4 cm
18 kg




104 cm
13.4 cm
21 kg




30 cm
7.6 cm
4.1 kg




42 cm
7.6 cm
4.4 kg

Optical Design Astigmatic two-mirror system with central coupling hole.
Mirrors Nickel plated aluminum with broad-band dielectric coating.
Wavelength range 3 to 11 µm (actual range may be broader).
Reflectivity 99.2 %
Window material BaF2
Body materials Nickel plated aluminum, stainless steel end flanges, Pyrex tube.

*Now available:  New Lightweight Astigmatic Multipass Absorption Cell

For more information about this product or associated R&D service contact:  

Dr. Mark Zahniser
Vice President
Director, Center for Atmospheric and Environmental Chemistry

Phone: 978-663-9500, Ext. 224
Fax: 978-663-4918
E-mail: mz@aerodyne.com

 

Mr. James Akimchuk, Manager of Adminstration

Phone: 978-663-9500, Ext. 214
Fax: 978-663-4918
E-mail: jima@aerodyne.com

Distributor in Japan:

Sanyu Plant Service Co., Ltd.
1-8-21 Hashimotodai, Sagamihara-shi
Kanagawa-ken 229-1132, Japan
Dr. Akio Shimono
Director, Equipment Div.
Tel: +81-427-73-3611
Fax: +81-427-74-1294
E-mail: A.Shimono@g-sanyu.co.jp

 

Patent:

"Off Axis Cavity Absorption Cell", #5,291,265, P.L. Kebabian, (Mar. 1, 1994).

References:

"Paraxial Matrix Description of Astigmatic and Cylindrical Mirror Resonators with Twisted Axes for Laser Spectroscopy," B. McManus,  App. Optics, 46, 472-482, (2007).

"Astigmatic mirror multipass absorption cells for long-path-length spectroscopy," J.B. McManus, P.L. Kebabian, and M.S. Zahniser, Applied Optics, 34, 3336-3348 (1995). (Abstract)

"Measurement of Trace Gas Fluxes Using Tunable Diode Laser Spectroscopy," M.S. Zahniser, D.D. Nelson, J.B. McManus and P.L. Kebabian. Phil. Trans. Roy. Soc. Lond. A., 351, 371-382 (1995).

 

Aerodyne Research, Inc.  45 Manning Road Billerica, MA  01821-3976  978 663-9500